MM40 - HP UNIT

Technical Overview

MM40 HP unit is working on the principal of different pressure between the operation and reserve regulator. By the manual lever, the operator can decide which side will be the operational side and which will be the reserve side. Manifolds are supplied with an Alarm system which increases safety to maximal level and informs the hospital personal about each non standard situation.
Gas Alarm is a standard accessory
The gas alarm gives visual and audible indications
The gas alarm is able to communicate with other equipment through relays
Manifold MM40 is only first stage of regulation and must be installed together with a Stabilizer which will stabilize the final pressure used in the hospital gas outlets
GCE medical manifolds are CE-marked and fulfill the ISO 7396-1 standard

Product Information

Item No.
Description
0732307
Denomination: MM40 – HP unit 2×1, Gas: O2, Air, N2, PRV: Standard, Signalization: Alarm TMA6703
0732308
Denomination: MM40 – HP unit 2×1, Gas: O2, Air, N2, PRV: Manual activation, Signalization: Alarm TMA6703
0727334*
Denomination: MM40 – HP unit 2×1, Gas: O2, Air, N2, N2O, CO2, PRV: Standard
0732310
Denomination: MM40 – HP unit 2×1, Gas: N2O, CO2, PRV: Standard, Signalization: Alarm TMA6703
0732311
Denomination: MM40 – HP unit 2×1, Gas: N2O, CO2, PRV: Manual activation, Signalization Alarm TMA6703
0732307
Denomination: MM40 – HP unit 2×1, Gas: O2, Air, N2, PRV: Standard, Signalization: Alarm TMA6703
0732308
Denomination: MM40 – HP unit 2×1, Gas: O2, Air, N2, PRV: Manual activation, Signalization: Alarm TMA6703
0727334*
Denomination: MM40 – HP unit 2×1, Gas: O2, Air, N2, N2O, CO2, PRV: Standard
0732310
Denomination: MM40 – HP unit 2×1, Gas: N2O, CO2, PRV: Standard, Signalization: Alarm TMA6703
0732311
Denomination: MM40 – HP unit 2×1, Gas: N2O, CO2, PRV: Manual activation, Signalization Alarm TMA6703

Technical

Gases:
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow
40 m3/h
Inlet nominal pressure
200 bar
Outlet nominal pressure
12 bar (setting range 9–16 bar)
Inlet connection
W21,8×1/14"M
Outlet connection
G1/2"M + soldering piece pipe ø 10, ø 15 mm
Pressure relieve valve
17 bar
Pressure relieve valve pipe dimension
ø 10 mm
Regulatory status
Complies with Medical Devices Directive 93/42/EEC, EN ISO 7396-1 (Central Gas Supply Systems), EN 60601-1-2 (Electromagnetic compatibility), present SIS HB 370 and HTM 02-01
Gases:
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow
40 m3/h
Inlet nominal pressure
200 bar
Outlet nominal pressure
12 bar (setting range 9–16 bar)
Inlet connection
W21,8×1/14"M
Outlet connection
G1/2"M + soldering piece pipe ø 10, ø 15 mm
Pressure relieve valve
17 bar
Pressure relieve valve pipe dimension
ø 10 mm
Regulatory status
Complies with Medical Devices Directive 93/42/EEC, EN ISO 7396-1 (Central Gas Supply Systems), EN 60601-1-2 (Electromagnetic compatibility), present SIS HB 370 and HTM 02-01