MM40 - Stabilizer

Technical Overview

Second stage pressure reduction unit with the task to equalize the eventual pressure variation in the hospital pipeline system to ensure a correct pressure from the terminal units. The stabilizer can be delivered with either pressure transmitter 4–20 mA, pressure transmitter 0–50 mV or with pressure switches.
Gas reduction unit MM40 – STABILIZER must always be installed in compliance with the standards EN ISO 7396-1 and the appropriate national standards
In case of signal for pressure deviation in relation to the alarm settings, the alarm can easily be displayed on a Gas alarm unit.
It is also possible to send information to the central operation control.

Product Information

Item No.
Description
0727329
Denomination: MM40 – Stabilizer, Gas: O2, N2O, Air, CO2, N2
0727337
Denomination: MM40 – Stabilizer, Gas: Air–800
0727332
Denomination: MM40 – Stabilizer, Gas: O2, N2O, Air, CO2, N2, PRV: Manual Activation, Signalization: Pressure switch
0727344
Denomination: MM40 – Stabilizer, Gas: O2, N2O, Air, CO2, N2, PRV: Manual Activation, Signalization: 4-20mA
0727343
Denomination: MM40 – Stabilizer, Gas: O2
0727346
Denomination: MM40 – Stabilizer, Gas: Air
0727347
Denomination: MM40 – Stabilizer, Gas: Air–800
0727348
Denomination: MM40 – Stabilizer, Gas: CO2
0727364
Denomination: MM40 – Stabilizer, Gas: N2O
0727329
Denomination: MM40 – Stabilizer, Gas: O2, N2O, Air, CO2, N2
0727337
Denomination: MM40 – Stabilizer, Gas: Air–800
0727332
Denomination: MM40 – Stabilizer, Gas: O2, N2O, Air, CO2, N2, PRV: Manual Activation, Signalization: Pressure switch
0727344
Denomination: MM40 – Stabilizer, Gas: O2, N2O, Air, CO2, N2, PRV: Manual Activation, Signalization: 4-20mA
0727343
Denomination: MM40 – Stabilizer, Gas: O2
0727346
Denomination: MM40 – Stabilizer, Gas: Air
0727347
Denomination: MM40 – Stabilizer, Gas: Air–800
0727348
Denomination: MM40 – Stabilizer, Gas: CO2
0727364
Denomination: MM40 – Stabilizer, Gas: N2O

Technical

Gases
O2, Air, Air–800, N2, N2O, CO2 (all medical gases)
Nominal flow
40 m3/h
Inlet maximal pressure
16 bar
Outlet nominal pressure
4,5 bar (setting range 0,5–10 bar)
Inlet connection 1
G1/2"M + soldering piece pipe ø 12 mm
Inlet connection 2
Optional (G1/2"M + soldering piece pipe ø 12 mm)
Outlet connection
G1/2"M + soldering piece pipe ø 12 mm
Pressure relieve valve
Optional (6,8 bar; outlet pipe ø 15 mm)
Pressure sensors
Optional (Pressure switches; Transmitter 4–20 mA)
Emergency QC inlet
Optional QC by national standards
Regulatory status
Complies with Medical Devices Directive 93/42/EEC, EN ISO 7396-1 (Central Gas Supply Systems), EN 60601-1-2 (Electromagnetic compatibility), present SIS HB 370 and HTM 02-01
Gases
O2, Air, Air–800, N2, N2O, CO2 (all medical gases)
Nominal flow
40 m3/h
Inlet maximal pressure
16 bar
Outlet nominal pressure
4,5 bar (setting range 0,5–10 bar)
Inlet connection 1
G1/2"M + soldering piece pipe ø 12 mm
Inlet connection 2
Optional (G1/2"M + soldering piece pipe ø 12 mm)
Outlet connection
G1/2"M + soldering piece pipe ø 12 mm
Pressure relieve valve
Optional (6,8 bar; outlet pipe ø 15 mm)
Pressure sensors
Optional (Pressure switches; Transmitter 4–20 mA)
Emergency QC inlet
Optional QC by national standards
Regulatory status
Complies with Medical Devices Directive 93/42/EEC, EN ISO 7396-1 (Central Gas Supply Systems), EN 60601-1-2 (Electromagnetic compatibility), present SIS HB 370 and HTM 02-01

Accessories

Item No.
Description
COM001002
Denomination: Lockable cover
COM001002
Denomination: Lockable cover

Technical Downloads