Manifold MM40 – STABILIZER is a second stage pressure reduction unit with the task to equalize the eventual pressure variation in the hospital pipeline system to ensure a correct pressure from the terminal units.
MM40 – STABILIZER is only a second stage reduction unit where the primary gas supply is provided by high pressure gas manifolds (such as MM40 – HP Unit). In case of a signal for pressure deviation in relation
to the alarm settings, the alarm can easily be displayed on a gas alarm unit. It is also possible to send information to the central operation control. The stabilizer can be delivered with either pressure transmitter
4–20 mA or with pressure switches. Gas reduction unit MM40 – STABILIZER must always be installed in compliance with the standards EN ISO 7396-1 and the appropriate national standards.
FOR A COMPLETE MM40 STABILIZER MANIFOLD ADD:
Plastic cover for locking
Alarm unit (included if ordered together with HP unit)
O2, Air, Air–800, N2, N2O, CO2 (all medical gases)
Nominal flow
40 m3/h
Inlet maximal pressure
16 bar
Outlet nominal pressure
4,5 bar (setting range 0,5–10 bar)
Inlet connection 1
G1/2"M + soldering piece pipe ø 12 mm
Inlet connection 2
Optional (G1/2"M + soldering piece pipe ø 12 mm)
Outlet connection
G1/2"M + soldering piece pipe ø 12 mm
Pressure relieve valve
Optional (6,8 bar; outlet pipe ø 15 mm)
Pressure sensors
Optional (Pressure switches; Transmitter 4–20 mA)
Emergency QC inlet
Optional QC by national standards
Regulatory status
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
Gases
O2, Air, Air–800, N2, N2O, CO2 (all medical gases)
Nominal flow
40 m3/h
Inlet maximal pressure
16 bar
Outlet nominal pressure
4,5 bar (setting range 0,5–10 bar)
Inlet connection 1
G1/2"M + soldering piece pipe ø 12 mm
Inlet connection 2
Optional (G1/2"M + soldering piece pipe ø 12 mm)
Outlet connection
G1/2"M + soldering piece pipe ø 12 mm
Pressure relieve valve
Optional (6,8 bar; outlet pipe ø 15 mm)
Pressure sensors
Optional (Pressure switches; Transmitter 4–20 mA)
Emergency QC inlet
Optional QC by national standards
Regulatory status
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
We use cookies on our website to give you the most relevant experience by remembering your preferences and repeat visits. By clicking “Accept”, you consent to the use of ALL the cookies.
This website uses cookies to improve your experience while you navigate through the website. Out of these, the cookies that are categorized as necessary are stored on your browser as they are essential for the working of basic functionalities of the website. We also use third-party cookies that help us analyze and understand how you use this website. These cookies will be stored in your browser only with your consent. You also have the option to opt-out of these cookies. But opting out of some of these cookies may affect your browsing experience.