Line Regulator

Capacity of 40 m3/h

Technical Overview

A stabilizer is a pressure reduction unit with the task to equalize the eventual pressure variation in the hospital pipeline system to ensure a correct pressure from the terminal units. The stabilizer makes it possible to distribute gas with different pressure to department and buildings in the hospital area.

Product Information

Item No.
Description
0727333
Denomination: LINE REG, Gas: O2, N2O, Air, CO2, N2, Inlet*: LH
K141621
Denomination: LINE REG, Gas: O2 – AFNOR, Inlet*: LH
K141631
Denomination: LINE REG, Gas: O2 – AFNOR, Inlet*: RH
K141622
Denomination: LINE REG, Gas: N2O – AFNOR, Inlet*: LH
K141632
Denomination: LINE REG, Gas: N2O – AFNOR, Inlet*: RH
K141623
Denomination: LINE REG, Gas: Air – AFNOR, Inlet*: LH
K141633
Denomination: LINE REG, Gas: Air – AFNOR, Inlet*: RH
K141629
Denomination: LINE REG, Gas: Air-800 – AFNOR, Inlet*: LH
K141639
Denomination: LINE REG, Gas: Air-800 – AFNOR, Inlet*: RH
K141624
Denomination: LINE REG, Gas: N2 – AFNOR, Inlet*: LH
K141625
Denomination: LINE REG, Gas: CO2 – AFNOR, Inlet*: LH
*LH = inlet from left side; RH = inlet from right side
0727333
Denomination: LINE REG, Gas: O2, N2O, Air, CO2, N2, Inlet*: LH
K141621
Denomination: LINE REG, Gas: O2 – AFNOR, Inlet*: LH
K141631
Denomination: LINE REG, Gas: O2 – AFNOR, Inlet*: RH
K141622
Denomination: LINE REG, Gas: N2O – AFNOR, Inlet*: LH
K141632
Denomination: LINE REG, Gas: N2O – AFNOR, Inlet*: RH
K141623
Denomination: LINE REG, Gas: Air – AFNOR, Inlet*: LH
K141633
Denomination: LINE REG, Gas: Air – AFNOR, Inlet*: RH
K141629
Denomination: LINE REG, Gas: Air-800 – AFNOR, Inlet*: LH
K141639
Denomination: LINE REG, Gas: Air-800 – AFNOR, Inlet*: RH
K141624
Denomination: LINE REG, Gas: N2 – AFNOR, Inlet*: LH
K141625
Denomination: LINE REG, Gas: CO2 – AFNOR, Inlet*: LH
*LH = inlet from left side; RH = inlet from right side
Regional information available in your language at
GCEGROUP.COM

Technical

Gases
O2, Air, Air–800, N2, N2O, CO2 (all medical gases)
Nominal flow
40 m3/h
Inlet nominal pressure
16 bar (1600 kPa)
Outlet nominal pressure
4,5 bar (setting range 0,5–10 bar)
Inlet connection
G1/2"M + soldering piece pipe ø 12 mm
Outlet connection
G1/2"M + soldering piece pipe ø 12 mm
Pressure sensors
Optional (Pressure switches; Transmitter 4–20 mA)
Emergency QC inlet
Optional QC by national standards
Regulatory status
Complies with Medical Devices Directive 93/42/EEC, EN ISO 7396-1 (Central Gas Supply Systems), EN 60601-1-2 (Electromagnetic compatibility), present HTM 02-01
Gases
O2, Air, Air–800, N2, N2O, CO2 (all medical gases)
Nominal flow
40 m3/h
Inlet nominal pressure
16 bar (1600 kPa)
Outlet nominal pressure
4,5 bar (setting range 0,5–10 bar)
Inlet connection
G1/2"M + soldering piece pipe ø 12 mm
Outlet connection
G1/2"M + soldering piece pipe ø 12 mm
Pressure sensors
Optional (Pressure switches; Transmitter 4–20 mA)
Emergency QC inlet
Optional QC by national standards
Regulatory status
Complies with Medical Devices Directive 93/42/EEC, EN ISO 7396-1 (Central Gas Supply Systems), EN 60601-1-2 (Electromagnetic compatibility), present HTM 02-01