Talk to the team

MC80 - DUPLEX

Capacity of 200 m3/h

Technical Overview

The MC80 gas manifold is suitable for medium to large sized hospital. It has a flow capacity up to 200 m3/h and is conveniently designed in modules. The MC80 reduces the gas pressure in two steps to a constant distribution pressure. Service and tests can be carried out with no disturbance in the supply of gas to the gas distribution system.
DUPLEX (MC80) INCLUDES THE FOLLOWING COMPONENTS:
MC80 – HP Unit
MC80 – Stabilizer
Alarm or just sensor based on item in column "Alarm"
Evacuating kits for collecting pipe
Shut-off valve for the distribution line
HP filters

Product Information

Item No.
Description
0732328
Denomination: DUPLEX 2×1, Gas: O2, RPV: Manual activation, Signalization: Alarm TMA9705
0732329
Denomination: DUPLEX 2×1, Gas: Air, RPV: Manual activation, Signalization: Alarm TMA9705
0732330
Denomination: DUPLEX 2×1, Gas: N2O/CO2, RPV: Manual activation, Signalization: Alarm TMA9705
0732328
Denomination: DUPLEX 2×1, Gas: O2, RPV: Manual activation, Signalization: Alarm TMA9705
0732329
Denomination: DUPLEX 2×1, Gas: Air, RPV: Manual activation, Signalization: Alarm TMA9705
0732330
Denomination: DUPLEX 2×1, Gas: N2O/CO2, RPV: Manual activation, Signalization: Alarm TMA9705
Regional information available in your language at
GCEGROUP.COM

Technical

Gases
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow
200 m3/h
Inlet maximal pressure
20 bar (2000 kPa)
Outlet nominal pressure
4,5 bar (setting range 0,5–6 bar)
Inlet connection
G3/4"F
Outlet connection
G3/4"F
Pressure relieve valve
6,8 bar
Pressure relieve valve pipe dimension
ø 15 mm
Regulatory status
Complies with Medical Devices Directive 93/42/EEC, EN ISO 7396-1 (Central Gas Supply Systems), EN 60601-1-2 (Electromagnetic compatibility), present SIS HB 370 and HTM 02-01
Gases
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow
200 m3/h
Inlet maximal pressure
20 bar (2000 kPa)
Outlet nominal pressure
4,5 bar (setting range 0,5–6 bar)
Inlet connection
G3/4"F
Outlet connection
G3/4"F
Pressure relieve valve
6,8 bar
Pressure relieve valve pipe dimension
ø 15 mm
Regulatory status
Complies with Medical Devices Directive 93/42/EEC, EN ISO 7396-1 (Central Gas Supply Systems), EN 60601-1-2 (Electromagnetic compatibility), present SIS HB 370 and HTM 02-01