MC80 - STABILIZER

Capacity of 200 m3/h

Technical Overview

Second stage pressure reduction unit with the task to equalize the eventual pressure variation in the hospital pipeline system to ensure a correct pressure from the terminal units. MC80 – STABILIZER is only second stage reduction unit where the primary gas supply is provided by high pressure gas manifolds. The alarm can easily be displayed on a Gas alarm unit.
It is also possible to send information to the central operation control
The stabilizer can be delivered with either pressure transmitter 4–20 mA or with contact gauge
Gas reduction unit MC80 – STABILIZER must always be installed in compliance with the standards EN ISO 7396-1 and the appropriate national standards

Product Information

Item No.
Description
0727324
Denomination: MC80 STAB, Gas: O2, RPV: Manual activation, Signalization: 4–20 mA
0727325
Denomination: MC80 STAB, Gas: Air, RPV: Manual activation, Signalization: 4–20 mA
0727326
Denomination: MC80 STAB, Gas: N2O, CO2, RPV: Manual activation, Signalization: 4–20 mA
0727339
Denomination: MC80 STAB, Gas: ALL, RPV: Manual activation, Signalization: Contact gauge
0727340
Denomination: MC80 STAB, Gas: ALL, RPV: Manual activation, Signalization: 4–20 mA
0727324
Denomination: MC80 STAB, Gas: O2, RPV: Manual activation, Signalization: 4–20 mA
0727325
Denomination: MC80 STAB, Gas: Air, RPV: Manual activation, Signalization: 4–20 mA
0727326
Denomination: MC80 STAB, Gas: N2O, CO2, RPV: Manual activation, Signalization: 4–20 mA
0727339
Denomination: MC80 STAB, Gas: ALL, RPV: Manual activation, Signalization: Contact gauge
0727340
Denomination: MC80 STAB, Gas: ALL, RPV: Manual activation, Signalization: 4–20 mA
Regional information available in your language at
GCEGROUP.COM

Technical

Gases
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow
200 m3/h
Inlet maximal pressure
20 bar (2000 kPa)
Outlet nominal pressure
4,5 bar (setting range 0,5–6 bar)
Inlet connection
G3/4"F
Outlet connection
G3/4"F
Pressure relieve valve
6,8 bar
Pressure relieve valve pipe dimension
ø 15 mm
Regulatory status
Complies with Medical Devices Directive 93/42/EEC, EN ISO 7396-1 (Central Gas Supply Systems), EN 60601-1-2 (Electromagnetic compatibility), present SIS HB 370 and HTM 02-01
Gases
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow
200 m3/h
Inlet maximal pressure
20 bar (2000 kPa)
Outlet nominal pressure
4,5 bar (setting range 0,5–6 bar)
Inlet connection
G3/4"F
Outlet connection
G3/4"F
Pressure relieve valve
6,8 bar
Pressure relieve valve pipe dimension
ø 15 mm
Regulatory status
Complies with Medical Devices Directive 93/42/EEC, EN ISO 7396-1 (Central Gas Supply Systems), EN 60601-1-2 (Electromagnetic compatibility), present SIS HB 370 and HTM 02-01