Talk to the team

MM40 - HP UNIT

Capacity of 40 m3/h

Technical Overview

Manifold MM40 HP unit is an automatic manifold. It is working on the principal of different pressures between the operation and reserve regulator. By the manual lever, the operator can decide which side will be the operational side and which will be the reserve side. When the operating side is empty, the manifold will without any action start to supply gas from the reserve side with the lower regulator pressure and fulfill the requests to supply without interrupting the flow.
Gas Alarm is a standard accessory
The gas alarm gives visual and audible indications
The gas alarm is able to communicate with other equipment through relays
Manifold MM40 is only first stage of regulation and must be installed together with a Stabilizer which will stabilize the final pressure used in the hospital gas outlets
GCE medical manifolds are CE-marked and fulfill the ISO 7396-1 standard

Product Information

Item No.
Description
0732307
Denomination: MM40 – HP unit 2×1
Gas: O2, Air, N2
PRV: Standard
Signalization: Alarm TMA6703
0732308
Denomination: MM40 – HP unit 2×1
Gas: O2, Air, N2
PRV: With pull-ring
Signalization: Alarm TMA6703
0727334*
Denomination: MM40 – HP unit 2×1
Gas: O2, Air, N2, N2O, CO2
PRV: Standard
*basic version without electric sensors
0732310
Denomination: MM40 – HP unit 2×1
Gas: N2O, CO2
PRV: Standard
Signalization: Alarm TMA6703
0732311
Denomination: MM40 – HP unit 2×1
Gas: N2O, CO2
PRV: With pull-ring
Signalization Alarm: TMA6703
0727371**
Denomination: MM40 – HP unit 2×1
Gas: O2, Air, N2
PRV: Standard
Signalization: 4-20 mA
**ISO standard version without Leaking reserve as requested for HB370
0732307
Denomination: MM40 – HP unit 2×1
Gas: O2, Air, N2
PRV: Standard
Signalization: Alarm TMA6703
0732308
Denomination: MM40 – HP unit 2×1
Gas: O2, Air, N2
PRV: With pull-ring
Signalization: Alarm TMA6703
0727334*
Denomination: MM40 – HP unit 2×1
Gas: O2, Air, N2, N2O, CO2
PRV: Standard
*basic version without electric sensors
0732310
Denomination: MM40 – HP unit 2×1
Gas: N2O, CO2
PRV: Standard
Signalization: Alarm TMA6703
0732311
Denomination: MM40 – HP unit 2×1
Gas: N2O, CO2
PRV: With pull-ring
Signalization Alarm: TMA6703
0727371**
Denomination: MM40 – HP unit 2×1
Gas: O2, Air, N2
PRV: Standard
Signalization: 4-20 mA
**ISO standard version without Leaking reserve as requested for HB370
Regional information available in your language at
GCEGROUP.COM

Technical

Gases:
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow
40 m3/h
Inlet nominal pressure
200 bar
Outlet nominal pressure
12 bar (setting range 9–16 bar)
Inlet connection
W21,8×1/14"M
Outlet connection
G1/2"M + soldering piece pipe ø 10, ø 15 mm
Pressure relieve valve
17 bar
Pressure relieve valve pipe dimension
ø 10 mm
Purge valves connection
W21,8×team1/14"M
Regulatory status
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
Gases:
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow
40 m3/h
Inlet nominal pressure
200 bar
Outlet nominal pressure
12 bar (setting range 9–16 bar)
Inlet connection
W21,8×1/14"M
Outlet connection
G1/2"M + soldering piece pipe ø 10, ø 15 mm
Pressure relieve valve
17 bar
Pressure relieve valve pipe dimension
ø 10 mm
Purge valves connection
W21,8×team1/14"M
Regulatory status
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01