Talk to the team

MM90 - HP UNIT FULLY AUTOMATIC

Capacity of 90 m3/h

Technical Overview

The MM90 HP unit medical manifold is intended for use in hospital pipeline system as medical gas source. Together with MM90, always use an alarm providing all alarms according to standard (like gas alarm). As 2nd stage is recommended to use a stabilizer. The manifold will deliver gas from the operating bank to the manifold pressure regulator until the cylinders are exhausted. At that point the supply will switch to the reserve bank and the empty bank can be replenished. The object gives uninterrupted gas supply. Gas Alarm is a standard accessory. The gas alarm gives visual and audible indication.

It surveils and the alarm sounds when the following happens:
1. Change operation side / Leaking on reserve side
2. High operation pressure
3. Low operation pressure
4. Empty possision (High/Low distribution pressure when connected to a Stabilizer)

The Gas alarm is able to communicate with other equipment through relays.
MM90 INCLUDES THE FOLLOWING COMPONENTS: MM90 HP unit Manifold, Gas alarm, Purge valves, HP filtres
FOR A COMPLETE MM90 HP UNIT MANIFOLD ADD: collecting pipe set (high pressure valves and non-return valves/high pressure components), cylinder retaining brackets (included in gas cylinder collecting pipe set), high pressure hoses with safety wire, plug for close collecting pipeline, gas name sign, stabilizer

Gallery

Product Information

Item No.
Description
0732322
Denomination: MM90 – HP unit AUTO 2×1
Gas: O2
PRV: Standard
Signalization: Alarm TMA6703
0732323
Denomination: MM90 – HP unit AUTO 2×1
Gas: Air
PRV: Standard
Signalization: Alarm TMA6703
0732324
Denomination: MM90 – HP unit AUTO 2×1
Gas: N2O, CO2
PRV: Standard
Signalization: Alarm TMA6703
0727308*
Denomination: MM90 – HP unit AUTO 2×1
Gas: O2, N2O, Air, CO2, N2
PRV: Standard
0727309
Denomination: MM90 – HP unit AUTO 2×1
Gas: O2
PRV: Standard
Signalization: Pressure switch
0727310
Denomination: MM90 – HP unit AUTO 2×1
Gas: Air
PRV: Standard
Signalization: Pressure switch
0727311
Denomination: MM90 – HP unit AUTO 2×1
Gas: N2O, CO2
PRV: Standard
Signalization: Pressure switch
*basic version without electric sensors
0732322
Denomination: MM90 – HP unit AUTO 2×1
Gas: O2
PRV: Standard
Signalization: Alarm TMA6703
0732323
Denomination: MM90 – HP unit AUTO 2×1
Gas: Air
PRV: Standard
Signalization: Alarm TMA6703
0732324
Denomination: MM90 – HP unit AUTO 2×1
Gas: N2O, CO2
PRV: Standard
Signalization: Alarm TMA6703
0727308*
Denomination: MM90 – HP unit AUTO 2×1
Gas: O2, N2O, Air, CO2, N2
PRV: Standard
0727309
Denomination: MM90 – HP unit AUTO 2×1
Gas: O2
PRV: Standard
Signalization: Pressure switch
0727310
Denomination: MM90 – HP unit AUTO 2×1
Gas: Air
PRV: Standard
Signalization: Pressure switch
0727311
Denomination: MM90 – HP unit AUTO 2×1
Gas: N2O, CO2
PRV: Standard
Signalization: Pressure switch
*basic version without electric sensors
Regional information available in your language at
GCEGROUP.COM

Technical

Gases:
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow:
90 m3/h
Inlet nominal pressure:
200 bar (20 000 kPa)
Outlet nominal pressure:
9 bar (setting range 9 – 15 bar)
Inlet connection:
W21,8×1/14"M
Outlet connection:
G1/2"M + soldering piece pipe ø 22 mm
Intermediate pressure relieve valve:
16 bar
Pressure relieve valve pipe dimension:
ø 10 mm
Purge valves connection
W21,8×1/14"M + soldering piece pipe ø 10 mm
Regulatory status:
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
Gases:
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow:
90 m3/h
Inlet nominal pressure:
200 bar (20 000 kPa)
Outlet nominal pressure:
9 bar (setting range 9 – 15 bar)
Inlet connection:
W21,8×1/14"M
Outlet connection:
G1/2"M + soldering piece pipe ø 22 mm
Intermediate pressure relieve valve:
16 bar
Pressure relieve valve pipe dimension:
ø 10 mm
Purge valves connection
W21,8×1/14"M + soldering piece pipe ø 10 mm
Regulatory status:
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01

Technical Downloads

Find a distributor

linkedin facebook pinterest youtube rss twitter instagram facebook-blank rss-blank linkedin-blank pinterest youtube twitter instagram