Talk to the team

MM90 - HP UNIT SEMI AUTOMATIC

Capacity of 90 m3/h

Technical Overview

The MM90 HP unit medical manifold is intended for use in hospital pipeline system as medical gas source. Together with MM90, always use an alarm providing all alarms according to standard (like gas alarm).
Gas Alarm is a standard accessory
The gas alarm gives visual and audible indication
The gas alarm is able to communicate with other equipment through relays
The manifold will deliver gas from the operating bank to the manifold pressure regulator until the cylinders are exhausted
At that point the supply will switch to the reserve bank and the exhausted bank can be replenished
The object gives uninterrupted gas supply

Product Information

Item No.
Description
0732314
Denomination: MM90 – HP unit 2×1, Gas: O2, RPV: Standard, Signalization: Alarm TMA6703
0732315
Denomination: MM90 – HP unit 2×1, Gas: Air, RPV: Standard, Signalization: Alarm TMA6703
0732316
Denomination: MM90 – HP unit 2×1, Gas: N2O, CO2, RPV: Standard, Signalization: Alarm TMA6703
0732320
Denomination: MM90 – HP unit 2×1, Gas: O2, Air, RPV: Manual activation, Signalization: Alarm TMA9705
0732321
Denomination: MM90 – HP unit 2×1, Gas: N2O, CO2, RPV: Manual activation, Signalization: Alarm TMA9705
0727327*
Denomination: MM90 – HP unit 2×1, Gas: O2, N2O, Air, CO2, N2, RPV: Standard
*basic version without electric sensors
0732314
Denomination: MM90 – HP unit 2×1, Gas: O2, RPV: Standard, Signalization: Alarm TMA6703
0732315
Denomination: MM90 – HP unit 2×1, Gas: Air, RPV: Standard, Signalization: Alarm TMA6703
0732316
Denomination: MM90 – HP unit 2×1, Gas: N2O, CO2, RPV: Standard, Signalization: Alarm TMA6703
0732320
Denomination: MM90 – HP unit 2×1, Gas: O2, Air, RPV: Manual activation, Signalization: Alarm TMA9705
0732321
Denomination: MM90 – HP unit 2×1, Gas: N2O, CO2, RPV: Manual activation, Signalization: Alarm TMA9705
0727327*
Denomination: MM90 – HP unit 2×1, Gas: O2, N2O, Air, CO2, N2, RPV: Standard
*basic version without electric sensors
Regional information available in your language at
GCEGROUP.COM

Technical

Gases
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow
90 m3/h
Inlet nominal pressure
200 bar (20 000 kPa)
Outlet nominal pressure
9 bar (setting range 9–15 bar)
Inlet connection
W21,8×1/14"M
Outlet connection
G3/4"F + soldering piece pipe ø 22 mm
Pressure relieve valve
16 bar
Pressure relieve valve dimension
ø 10 mm
Purge valves connection
W21,8×1/14"M + soldering piece pipe ø 10mm
Regulatory status
Complies with Medical Devices Directive 93/42/EEC, EN ISO 7396-1 (Central Gas Supply Systems), EN 60601-1-2 (Electromagnetic compatibility), present SIS HB 370 and HTM 02-01
Gases
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow
90 m3/h
Inlet nominal pressure
200 bar (20 000 kPa)
Outlet nominal pressure
9 bar (setting range 9–15 bar)
Inlet connection
W21,8×1/14"M
Outlet connection
G3/4"F + soldering piece pipe ø 22 mm
Pressure relieve valve
16 bar
Pressure relieve valve dimension
ø 10 mm
Purge valves connection
W21,8×1/14"M + soldering piece pipe ø 10mm
Regulatory status
Complies with Medical Devices Directive 93/42/EEC, EN ISO 7396-1 (Central Gas Supply Systems), EN 60601-1-2 (Electromagnetic compatibility), present SIS HB 370 and HTM 02-01

Technical Downloads

Find a distributor

GCE Headquarters
Källvattengatan 9, 212 23 Malmö, Sweden
1 2 3 174
linkedin facebook pinterest youtube rss twitter instagram facebook-blank rss-blank linkedin-blank pinterest youtube twitter instagram