MM90 - STANDBY BACKUP

Capacity of 90 m3/h

Technical Overview

The manifold MM90 STANDBY is designed to be used as a third source of supply in medical central gas systems. The manifold will deliver gas when the nominal supply system pressure falls below a set level (7 bar). Together with MM90 STANDBY, always use the MM90 HP unit and alarm providing all alarms according to standard. As 2nd stage is recommended to use a stabilizer.
Gas Alarm is a standard accessory
The Gas alarm gives visual and audible indication
It surveils and the alarm sounds when the following happens:
1. Too high outlet pressure
2. Too low outlet pressure
3. Empty cylinder
MM90 standby includes the following components:
MM90 standby manifold with pressure switches
Purge valves
HP filters

Product Information

Item No.
Description
0727307
Denomination: MM90 STANDBY, Gas: O2, Air, RPV: Standard, Signalization: Pressure switches
0727312
Denomination: MM90 STANDBY, Gas: N2O, CO2, RPV: Standard, Signalization: Pressure switches
0727338
Denomination: MM90 STANDBY, Gas: N2O, CO2, RPV: Manual activation, Signalization: Pressure switches
0727307
Denomination: MM90 STANDBY, Gas: O2, Air, RPV: Standard, Signalization: Pressure switches
0727312
Denomination: MM90 STANDBY, Gas: N2O, CO2, RPV: Standard, Signalization: Pressure switches
0727338
Denomination: MM90 STANDBY, Gas: N2O, CO2, RPV: Manual activation, Signalization: Pressure switches
Regional information available in your language at
GCEGROUP.COM

Technical

Gases
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow
90 m3/h
Inlet nominal pressure
200 bar (20 000 kPa)
Outlet nominal pressure
7 bar (setting range 7–15 bar)
Inlet connection
W21,8×1/14"M
Outlet connection
G3/4"F + soldering piece pipe ø 22 mm
Pressure relive valve
16 bar
Pressure relive valve pipe dimension
ø 10 mm
Purge valves connection
W21,8×1/14"M + soldering piece pipe ø 10 mm
Regulatory status
Complies with Medical Devices Directive 93/42/EEC, EN ISO 7396-1 (Central Gas Supply Systems), EN 60601-1-2 (Electromagnetic compatibility), present SIS HB 370 and HTM 02-01
Gases
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow
90 m3/h
Inlet nominal pressure
200 bar (20 000 kPa)
Outlet nominal pressure
7 bar (setting range 7–15 bar)
Inlet connection
W21,8×1/14"M
Outlet connection
G3/4"F + soldering piece pipe ø 22 mm
Pressure relive valve
16 bar
Pressure relive valve pipe dimension
ø 10 mm
Purge valves connection
W21,8×1/14"M + soldering piece pipe ø 10 mm
Regulatory status
Complies with Medical Devices Directive 93/42/EEC, EN ISO 7396-1 (Central Gas Supply Systems), EN 60601-1-2 (Electromagnetic compatibility), present SIS HB 370 and HTM 02-01