MM90 - STANDBY BACKUP

Capacity of 90 m3/h

Technical Overview

The manifold MM90 STANDBY is designed to be used as a third source of supply in medical central gas systems. The manifold will deliver gas when the nominal supply system pressure falls below a set level (7 bar). This is a back up source. Together with MM90 STANDBY always use the MM90 HP unit and alarm providing all alarms according to standard (like Gas alarm). As 2nd stage it is recommended to use a stabilizer. Gas Alarm is a standard accessory.

The Gas alarm gives visual and audible indication. It surveils and the alarm sounds when the following happens:
1. Too high outlet pressure
2. Too low outlet pressure
3. Empty cylinder
MM90 STANDBY INCLUDES THE FOLLOWING COMPONENTS: MM90 STANDBY Manifold with pressure switches, Purge valves, HP filters
FOR A COMPLETE MM90 STANDBY MANIFOLD ADD:
Gas alarm
Collecting pipe set (high pressure valves, and non-return valves / high pressure components)
Cylinder retaining brackets (included in gas cylinder collecting pipe set)
High pressure hoses with safety wire
Plug for close collecting pipeline
Gas name sign
Stabilizer

Product Information

Item No.
Description
0727307
Denomination: MM90 STANDBY
Gas: O2, Air
PRV: Standard
Signalization: Pressure switches
0727312
Denomination: MM90 STANDBY
Gas: N2O, CO2
PRV: Standard
Signalization: Pressure switches
0727338
Denomination: MM90 STANDBY
Gas: N2O, CO2
PRV: With pull-ring
Signalization: Pressure switches
0727367
Denomination: MM90 STANDBY
Gas: O2, Air
PRV: Standard
Signalization: 4-20mA pressure transmitter
0727368
Denomination: MM90 STANDBY
Gas: N2O, CO2
PRV: Standard
Signalization: 4-20mA pressure transmitter
0727307
Denomination: MM90 STANDBY
Gas: O2, Air
PRV: Standard
Signalization: Pressure switches
0727312
Denomination: MM90 STANDBY
Gas: N2O, CO2
PRV: Standard
Signalization: Pressure switches
0727338
Denomination: MM90 STANDBY
Gas: N2O, CO2
PRV: With pull-ring
Signalization: Pressure switches
0727367
Denomination: MM90 STANDBY
Gas: O2, Air
PRV: Standard
Signalization: 4-20mA pressure transmitter
0727368
Denomination: MM90 STANDBY
Gas: N2O, CO2
PRV: Standard
Signalization: 4-20mA pressure transmitter
Regional information available in your language at
GCEGROUP.COM

Technical

Gases:
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow:
90 m3/h
Inlet nominal pressure:
200 bar (20 000 kPa)
Outlet nominal pressure:
7 bar (setting range 7–15 bar)
Inlet connection:
W21,8×1/14"M
Outlet connection:
G3/4"F + soldering piece pipe ø 22 mm
Pressure relive valve:
16 bar
Pressure relive valve pipe dimension:
ø 10 mm
Purge valves connection:
W21,8×1/14"M + soldering piece pipe ø 10 mm
Regulatory status:
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
Gases:
O2, Air, N2, N2O, CO2 (all medical gases)
Nominal flow:
90 m3/h
Inlet nominal pressure:
200 bar (20 000 kPa)
Outlet nominal pressure:
7 bar (setting range 7–15 bar)
Inlet connection:
W21,8×1/14"M
Outlet connection:
G3/4"F + soldering piece pipe ø 22 mm
Pressure relive valve:
16 bar
Pressure relive valve pipe dimension:
ø 10 mm
Purge valves connection:
W21,8×1/14"M + soldering piece pipe ø 10 mm
Regulatory status:
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01

Technical Downloads

Find a distributor

linkedin facebook pinterest youtube rss twitter instagram facebook-blank rss-blank linkedin-blank pinterest youtube twitter instagram