Talk to the team

SIMPLEX MMR

Capacity of 30 m3/h

Technical Overview

The Simplex MMR gas manifold is suitable for such health care where the capacity requirement is limited, such as laboratories and small health care clinics, veterinary etc. This gas manifold consists of only one group of cylinders. The regulator is mounted in the collection unit.
Each inlet connection has a filter, a non-return valve and a shut-off valve. This arrangement makes it possible to use one cylinder at a time.
In order to obtain a stabile outlet pressure this gas manifold is equipped with a preset two-stage regulator.
On the high pressure side of the regulator there is a contact gauge the signal of which can be carried further to an alarm unit.
SIMPLEX MMR INCLUDES THE FOLLOWING COMPONENTS:
Gas cylinder manifold Simplex MMR
Collecting pipe many flow block for three hoses
Gas evacuation kits for collecting pipe

Product Information

Item No.
Description
325397702
Denomination: Simplex MMR, Gas: O2, Air, N2, Ar, N2O, CO2, RPV: Standard integrated, Signalization: Contact gauge
325397702
Denomination: Simplex MMR, Gas: O2, Air, N2, Ar, N2O, CO2, RPV: Standard integrated, Signalization: Contact gauge
Regional information available in your language at
GCEGROUP.COM

Technical

Gases
O2, Air, N2, Ar, N2O, CO2 (all medical gases)
Nominal flow
30 m3/h
Inlet nominal pressure
200 bar (20000 kPa)
Outlet nominal pressure
5 bar (setting range 4–5 bar)
Inlet connection
W21,8×1/14"M
Outlet connection
G3/8"M
Pressure relieve valve
6 bar
Pressure relieve valve pipe dimension
ø 8 mm
Purge valves connection
W21,8×1/14"M+pipe ø 15 mm
Regulatory status
Complies with Medical Devices Directive 93/42/EEC, EN ISO 7396-1 (Central Gas Supply Systems), EN 60601-1-2 (Electromagnetic compatibility), present SIS HB 370 and HTM 02-01
Gases
O2, Air, N2, Ar, N2O, CO2 (all medical gases)
Nominal flow
30 m3/h
Inlet nominal pressure
200 bar (20000 kPa)
Outlet nominal pressure
5 bar (setting range 4–5 bar)
Inlet connection
W21,8×1/14"M
Outlet connection
G3/8"M
Pressure relieve valve
6 bar
Pressure relieve valve pipe dimension
ø 8 mm
Purge valves connection
W21,8×1/14"M+pipe ø 15 mm
Regulatory status
Complies with Medical Devices Directive 93/42/EEC, EN ISO 7396-1 (Central Gas Supply Systems), EN 60601-1-2 (Electromagnetic compatibility), present SIS HB 370 and HTM 02-01